Vickers hardheidsmeter

De Vickers hardheidsmeter serie "Qness" brengt het Vickers hardheidsmeten op een geheel nieuw niveau. Onze high-end modellen van de laatste generatie combineren korte cyclustijden met maximale precisie. Eigenschappen zoals de 8-positie gereedschapswisselaar, uitgebreide software en connectiviteit via Qconnect maken deze Vickers hardheidsmeter serie de optimale keuze voor klanten die op zoek zijn naar de meest innovatieve producten op de markt.

Micro Vickers hardheidsmeter Testkracht bereik: 0.25 g – 62.5 kg

Macro Vickers hardheidsmeter Testkracht bereik: 0.3 kg – 3000 kg

Volledig Automatische Vickers hardheidsmeter Testkracht bereik: 0.3 kg – 3000 kg

.

Achtergrond informatie Vickers hardheidstest (DIN EN ISO 6507)

Based on the ideas of Smith and Sandland, a hardness testing method was developed in 1925 in the English Vickers works, using a diamond pyramid as the indenter. A regular four-sided diamond pyramid with a surface angle of 136° is pressed vertically into the preferably polished surface of the test specimen with a certain test load. After a certain exposure time, the test specimen is removed again and the two diagonals (d1 and d2) of the indentation are measured with a measuring microscope. The average length d is then calculated. The Vickers hardness is then determined according to the given formula.

1. Indentation of the diamond pyramid with test force F into the test specimen

Indentation of the diamond pyramid with test force F into the test specimen

2. Measurement of the indentation with diameter d

Measurement of the indentation with diameter d

Calculation of the Vickers hardness

Vickers calculation formula:

Calculation of the Vickers hardness

Prerequisites for standard-compliant testing

Various prerequisites need to be fulfilled for standard-compliant testing:

Prerequisites for standard-compliant testing

Test temperature

  • normal requirements: 10°C bis 35°C
  • high requirements: 23°C ± 5°C

Workpiece thickness

  • t ≥ 1.5 * d with t = sample or layer thickness d = average length of the indentation diagonals

Distances between the indentations and from the edge

  • For steels, Cu and Cu alloys
    • a ≥ 2.5  * d
    • b ≥ 3 * d
  • For light metals, Pb, Zn as well as their alloys
    • a ≥ 3 * d
    • b ≥ 6 * d

It is distinguished between the macro, small force and micro hardness range.

Macro Hardness Range

General tests on relatively thick components

Recommended test forces

Name of value

980.70

HV 100

490.30

HV 50

294.20

HV 30

196.10

HV 20

98.67

HV 10

49.03

HV 5

Small Force Range

Tests on thin workpieces, thin surface layers and foils

Recommended test forces

Name of value

29.42

HV 3

19.6

HV 2

9.807

HV 1

4.903

HV 0.5

2.942

HV 0.3

1.961

HV 0.2

Micro Hardness Range

Measurements at individual structure components

Recommended test forces

Name of value

0.981

HV 0.1

0.490

HV 0.05

0.245

HV 0.025

0.196

HV 0.02

0.098

HV 0.01

Vickers hardness test results

According to DIN EN ISO 6507-1, the result of a Vickers test is given as follows:

700 HV 10 / 20   ⇒   Hardness value as per Vickers

700 HV 10 / 20   ⇒   Abbreviation for Vickers hardness 

700 HV 10 / 20   ⇒   Test force specification: Test force [N] = 9.807 * test force specification

700 HV 10 / 20   ⇒   Application time of the test force (omitted if it is between 1-15 s)

measuring hardness depths

The Vickers test is also the preferred method for measuring hardness depths. A distinction is primarily made between the depth of hardening after case hardening (CHD), after nitriding (NHD) and after induction hardening (SHD).

Depth of hardening after case hardening (CHD)

When testing the case hardening depth (CHD), the hardness in HV1 is measured on the cross-section at regular distances from the surface. In order to comply with the above-mentioned rules of distance between the individual indentations, this is usually done as a 'zig-zag line'. The values are plotted graphically. The CHD is the distance at which the hardness falls below a certain limit. Usually, the limit is GH = 550 HV1 but a deviating value can be determined.

An example of this method is shown in the following image. The case hardening depth (CHD) can also be calculated from the pairs of values.

Distance from the surface [mm] ⇒ Hardness [HV1]

0.1 ⇒ 717
0.2 ⇒ 718
0.3 ⇒ 705
0.4 ⇒ 675
0.5 ⇒ 645
0.6 ⇒ 610
0.7 ⇒ 580
0.8 ⇒ 550
0.9 ⇒ 520
1.0 ⇒ 490
1.1 ⇒ 465
1.2 ⇒ 450
1.3 ⇒ 440
1.4 ⇒ 430
1.5 ⇒ 425

Distance from the surface [mm] ⇒ Hardness [HV1]

CHD = Case Hardening Depth

Depth of hardening after nitriding

When testing the nitriding hardening depth (NHD), the hardness in HV0.5 is measured on the cross-section at regular distances from the surface. In order to comply with the above-mentioned rules of distance between the individual indentations, this is usually done as a 'zig-zag line'. The values are plotted graphically. The NHD is the distance at which the hardness falls below a certain limit. Usually, the limit is defined as GH = core hardness + 50 HV, but a deviating value can be determined.

An example of this method is shown in the following image. The nitriding hardening depth (NHD) can also be calculated from the pairs of values.

Distance from the surface [mm] ⇒ Hardness [HV1]

0.05 ⇒ 1080
0.1 ⇒ 1050
0.2 ⇒ 350
0.3 ⇒ 257
0.4 ⇒ 250
0.5 ⇒ 250
0.6 ⇒ 250
0.7 ⇒ 250
0.8 ⇒ 250
0.9 ⇒ 250
1.0 ⇒ 250

Distance from the surface [mm] ⇒ Hardness [HV1]

NHD = Nitriding Hardening Depth

limit hardness GH = core hardness KH + 50 HV

Depth of hardening after induction hardening

When testing the induction hardening depth (SHD), the hardness in HV15 is measured on the cross-section at regular distances from the surface. In order to comply with the above-mentioned rules of distance between the individual indentations, this is usually done as a 'zig-zag line'. The values are plotted graphically. The SHD is the distance at which the hardness falls below a certain limit. Usually, the limit is defined as GH = 85% of the minimum surface hardness, but a deviating value can be determined.

An example of this method is shown in the following image. The induction hardening depth (SHD) can also be calculated from the pairs of values.

Distance from the surface [mm] ⇒ Hardness [HV1]

0.1 ⇒ 698
0.2 ⇒ 700
0.3 ⇒ 703
0.4 ⇒ 705
0.5 ⇒ 705
0.6 ⇒ 705
0.7 ⇒ 703
0.8 ⇒ 701
0.9 ⇒ 698
1.0 ⇒ 700
1.1 ⇒ 701
1.2 ⇒ 660
1.3 ⇒ 500
1.4 ⇒ 300
1.5 ⇒ 285
1.6 ⇒ 284
1.7 ⇒ 283
1.8 ⇒ 285
1.9 ⇒ 282
2.0 ⇒ 281
 

Distance from the surface [mm] ⇒ Hardness [HV1]

SHD = Surface Hardening Depth

Target58+4 HRC
650+100 HV10
⇒ GH85%(650HV)
⇒ GH =553 HV